Preoperative Planning Software Does Not Accurately Predict… : JAAOS – Journal of the American Academy of Orthopaedic Surgeons
Methods: This was a retrospective study of 16 reverse total shoulder arthroplasty patients with preoperative and postoperative (CT) scans obtained at least 1 year postoperatively. Clinical ROM was measured in abduction, external rotation at resting abduction, extension, and flexion at a minimum of 1 year postoperatively. All clinical measurements were obtained before generation of PPS ROM values. Using postoperative CT scans, the achieved implant component positions were quantified and then replicated in PPS on the preoperative CT scans. The preoperative predicted ROM was then recorded, both with and without osteophyte removal. Bland-Altman plots were generated within each motion comparing the differences between clinically measured motion and software-predicted motion. Results: The variation in clinically measured ROM in abduction, external rotation at resting abduction, extension, and flexion were 118 ± 27 (65° to 180°), 33 ± 16 (10° to 75°), 56 ± 8 (50° to 65°), and 137 ± 25 (80° to